This guide shows you how to conduct measurements with the Lilian device and the mobile app and store their results in the cloud.

Description of the measurement process

Step 1

Open the flap.

Step 2

Make sure to insert the SensoStick with the printed side up as demonstrated. Slide the SensoStick from back to front in one of the two slots.

Important:
Avoid touching the front of the SensoStick!

Step 3

Push the SensoStick deep into the slot until you hear an audible click takes place. You can close the lid now.

Step 4

Start the LILIAN Manager App to perform the measurement. The app guide you through the measurement process. Details on the handling can be found in the separate manual “Manual LILIAN Software”. You can download this at https://lilianlabs.com/downloads.

Step 5

Swipe over the touch sensor, while the app searches for the LILIAN. During the process, a white light moves along the LED ring. As soon as the smartphone has connected to the LILIAN, the LED ring turns white.

Step 6

Now rinse the measuring cup with clear water. Take your water sample afterwards. Make sure that the cup is filled as demonstrated in the picture. So the LILIAN can measure the sample in a correct way.

Important:
Avoid touching the water sample.

Step 7

Push the measuring cup with open lid into the recess of the measuring holder. Then place the LILIAN with the inserted SensoStick on the measuring holder as shown in the picture on the left. Now swipe the touch sensor of the LILIAN to start the measurement. If the device vibrates and lights up yellow, it is not correctly positioned in the measurement holder. In this case, please correct the position.

Step 8

While preparing and analyzing the water sample, a white light moves along the LED ring. After the LED ring glows green, you can pull the LILIAN out of the water sample.

Step 9

As soon as the led ring turns off, you can carefully pull the SensoStick out of the device from the front to remove it from the LILIAN. Please do not push it out of the device in the direction of the illuminated ring under any circumstances.

Step 10

A few seconds later, the measurement results are displayed on your smartphone. The data will be automatically transferred to the LILIAN Manager database. You need to be logged in for this to work.

Handling tips

Filling height

Use the measuring cup provided and make sure that the water level is correct. If the cup is not filled sufficiently, the LILIAN cannot pump the water sample correctly into the SensoStick, resulting in an error message.

The cup should be filled with the water sample up to the thread.

Handling the sample cup

Please do not touch the sample cup! Residues of your fingerprint cause the free chlorine in the water sample to react with the combined chlorine. This falsifies the measurement result.

This way, you hold the measuring cup just right.

Cleaning and storing cups

The measuring cup must be sufficiently rinsed and cleaned before and after each measurement. Do not store wet cups with the lid closed as germs can form. Store cups open and facing down on paper towels until dry. Replace measuring cups after one month of use at the latest.

We recommend storing the cups open and with the open side facing down, e.g. on kitchen roll, until they are dry. If you want to wipe the cups dry with a cloth, please take care not to get any particles or impurities into the cup. Replace the measuring cups after one month of use at the latest.

Opening SensoStick packs

The SensoStick packs have a notch that allows you to open the pack easily without scissors. Opening this way ensures you automatically grab the back of the SensoStick to avoid contamination.

This means that you automatically grab the back of the SensoStick to avoid this problem..

Touching the SensoStick

The part of the stick that is in the water sample must not be touched. The free chlorine of the water sample will other- wise react with traces of the fingerprint on the SensoStick and you will measure increased combined chlorine values.

The stick is only ever touched at the back when it is inserted into the measuring device in order to prevent deviations in the measurement result.

Insertion direction

In this example, the SensoStick is inserted from the wrong direction.

The SensoStick is always inserted into the opening of the LILIAN flap from behind.

Stick orientation

The SensoStick must not be inserted with the printed side down.

The SensoStick must not be inserted with the printed side down. Always insert into one of the two slots with the printed side facing up.

Removing the SensoStick

The SensoStick is always pulled out of the slots from back to front. Never push it through from front to back to avoid pushing the wet part back through the shaft.

Be careful not to remove the SensoStick from the shaft from front to back.

To avoid contamination, the LILIAN is designed in such a way that the wet part of the SensoStick does not have to be pushed through the shaft after a measurement.

Closing the flap

The flap of the handheld meter must be closed for each measurement to minimize the influence of stray light.

In this example you can see a correct measurement with the closed flap of the LILIAN.

Using the measurement holder

Never perform a measurement without the measuring holder. The LILIAN must remain in a vertical position during the entire measurement.

Here the LILIAN is placed in the special measuring holder. This has suitable recess and ensures that the LILIAN cannot change position during the measurement.

Cleaning the measurement shaft

Always ensure that the measuring shaft is clean. Particles, moisture or heavy scratches can falsify the measuring result.

The easiest way to clean the measuring channels is with a cotton pad moistened with isopropanol. During cleaning, perform gentle, circular movements in the measuring slot.

We also recommend sending the device to the manufacturer for maintenance once a year.

Seals and springs maintenance

The seals and springs in the LILIAN wear out over time and must be replaced regularly. Please replace them if they look like on the picture above.

The “legs” of the spring are pressed flat in the picture. You can find instructions about his topic on our website or contact us.

This is what functional springs and seals look like. Please check these elements as soon as the error message “No water sample could be taken.” is displayed to you for the first time.

App approvals

Activate all necessary app approvals for the system to function smoothly. For Android, these are location (exact), Bluetooth, memory sharing, allow notifications and nearby devices..

Important: Lilian Labs does not read location data or use approvals for marketing purposes. The requested permissions are used exclusively for the technical functionality of the app.

Charging safety during measurement

For safety reasons, the device cannot be charged during a measurement. Therefore, remove the charging plug during the measurement.

Here the measurement can take place because the charger is not plugged in. A minimum charge level of 10% is required to perform a measurement.

SensoSticks storage temparature

The storage temperature for the SensoSticks is 0-30 °C. However, we do not recommend storing them in the refrigerator, as they must first warm up to room temperature before they can be used. Otherwise, the SensoSticks will mist up and the measurement may be falsified.

Optimal storage for SensoSticks is at room temperature in an
environment that is not too humid. After opening the packaging, you should consume the SensoSticks within 7 days.

SensoSticks expiry date

The SensoSticks have a best-before date indicated on the outer packaging. After the expiry date, the sticks can usually still be used, but there may be minimal measurement deviations.

In order to avoid expired SensoSticks, we recommend to estimate the consumption accurately. Unfortunately, Lilian Labs cannot exchange expired SensoSticks.

Resetting the device

In very rare cases, the LILIAN measuring device may stop responding.

In this case, connect the device to the socket using the charging cable supplied. The software on the device will restart automatically.

If this does not help either, you must reset the LILIAN measuring device completely. To do this, bridge the two outer PINs of the 3-PIN combination on the device with a conductive object such as a paper clip, scissors or the springs from the maintenance kit. After approx. 1 second, the device vibrates and the reset is successful.

Diagnostic kit

The LILIAN water analysis system has been tested by independent institutes and laboratories and offers unprecedented accuracy in the field of mobile water analysis. If you still have reason to question the results measured with LILIAN because the measured values are unexpected and inappropriate you can use our diagnostic kit.

With the diagnostic kit, you can quickly and easily check whether your LILIAN measuring device is working correctly. You can quickly rule out a malfunction of the device.

Returning used SensoSticks and LILIAN devices

Please collect the used SensoSticks. We also take back old devices sold by us.

SensoSticks purchased directly from LilianLabs can be returned by mail. If you have purchased the system form one of our sales partners, please clarify with them how the return process is organized.

Lilian Labs GmbH will then dispose of the devices properly and the SesnoSticks will be sent for a quantity-dependent treatment or recycling process.

Charging the LILIAN

Check charge level

To check the charge level of the LILIAN, swipe the touch sensor without connecting via the app. If the LED ring is red, the battery level is less than 20%. Over 20%, the LED rings fills in 20% steps from 20% until 100%.

Charge the battery

The LILIAN must be recharged after about 100 measurements or 10 days in standby mode. To charge the LILIAN, plug the supplied charging cable into the charging socket of the switched- off device. The LED ring then shows a blue charging animation. Charging time is approx. 1 hour.

Additional info

Technical data

Device Type:
Integrated photometer for water analysis

Dimensions:
210mm x 88mm x 70mm (Weight: ~350g)

Battery:
2x Lithium polymer Akku (~100 measurements / 10 days standby)

Protection Class:
IP64 (splash-proof)

Connection:
Bluetooth

Charger:
In 230VAC / Out 24VDC